We are currently expanding our product catalogue and price data. Some pages may respond more slowly than usual over the next few days.
I lager hos oss 7 enheter redo att skickas från vårt eget lager — köp nu eller be oss om ett volympris.
701855 #VR45TQDPUN9W
Yokogawa Lägg ett bud

701855

Yokogawa 701855 High-Speed Isolation Module, 10 MS/s 12-bit, 1 ch, for DL708E / DL716

Kategori Other Automation
Enhet: pcs
Art.nr (normaliserat): 701855
I lager 7 (Begagnad)
Lägg ett bud
7 enheter i lager

Föreslå ditt eget pris för den här artikeln — vi svarar här.

Bud och priser är inte bindande förrän en beställning har lagts och bekräftats.

Tillgängliga enheter (7)

Om 701855

The Yokogawa 701855 (front-panel abbreviation HS_ISOL) is a single-channel, isolated voltage input module for Yokogawa's DL700-generation plug-in digital scopes: the 8-slot DL708E and the 16-slot DL716. Each module adds one isolated BNC input to the mainframe and digitizes it at up to 10 MS/s with 12-bit resolution.

The input is isolated and unbalanced, with AC, DC and GND coupling, 5 mV/div to 20 V/div sensitivity and a frequency response of DC to 1 MHz. A selectable bandwidth limit (500 Hz to 500 kHz) helps on noisy signals. Used directly it accepts 250 V (DC+AC peak); with Yokogawa's 700929 10:1 isolated probe the range extends to 850 V and 400 Vrms common mode, which suits inverter, motor-drive and power-line measurements where every channel must float independently.

It is not a DL750/DL850 module: those ScopeCorders use the later two-channel 7012xx/7202xx modules, such as the 701250. It is the 12-bit counterpart of the 10-bit 701850 used in the DL708/DL708E, and the DL716 manual notes that the DL708/DL708E modules 701850 and 701851 can also be used in the DL716.

Vanliga frågor

Does the 701855 fit a DL750 or DL850 ScopeCorder?

No. The 701855 belongs to the DL708E and DL716 digital scopes. The DL750 and DL850 series take the 7012xx/7202xx two-channel modules; the closest equivalent there is the 701250 high-speed 10 MS/s 12-bit isolation module.

Which probe gives the full 850 V rating?

Yokogawa's 700929 10:1 probe for isolated BNC inputs. Without it, the direct input is rated 250 V (DC+AC peak) and 42 V common mode.

Can the 701855 use the mainframe's external sample clock?

No. The DL716 manual lists the high-speed modules (701855, 701856 and the older 701850/701851) as not usable with the external clock input.

Dokument

Specifikationer

Input type Isolated unbalanced, BNC (isolated type)
DC accuracy +/-(1.5% of 8 div + offset accuracy) at 10 mV/div to 20 V/div
A/D resolution 12 bit (500 LSB/div)
Input channels 1
Input coupling AC, DC, GND
Bandwidth limit OFF / 500 Hz / 5 kHz / 50 kHz / 500 kHz
Input impedance 1 MOhm +/-1%, approx. 30 pF
Withstand voltage 3700 VAC for 1 minute, input to earth
Maximum sample rate 10 MS/s
Common mode rejection 80 dB or more at 50/60 Hz (typical)
Compatible mainframes Yokogawa DL708E, DL716
Maximum input voltage 250 V (DC+AC peak) direct; 850 V (DC+AC peak) with 700929 10:1 probe
Voltage axis sensitivity 5 mV/div to 20 V/div (50 mV/div to 200 V/div with 700929 probe)
Frequency response ( 3 dB) DC to 1 MHz
Maximum common mode voltage 42 V (DC+AC peak) direct; 400 Vrms with 700929 probe

Prishistorik Prisöversikt

Läser in …

Tillgängliga erbjudanden

Inga erbjudanden registrerade för den här produkten ännu.

Kommentarer

Inga kommentarer ännu — bli först.

Offert